Sibagropribor announces the sales release of a new model of high-speed milk analyzer developed specially to meet the requirements of the Indian market.
Milk quality analyzer "Laktan 1-4M" model 240 is used to determine the percentage of fat, solids-non-fat (SNF) in fresh whole, preserved, pasteurized, normalized, powder milk, non-fat and long storage milk. The analyzer may be used to conduct express analysis when preparing, receiving and processing milk as well as in selection work.
The main distinctintive feature of the new model is a combination of the high measurement speed (90 samples per hour), an affordable price and high reliability and accuracy.
Having a vast experience of work with Indian companies and in line with its strategy of reinforcing its positions in the Indian market Sibagropribor is now offering a new model of high-speed milk analyzer which fully meets the requirements of the Indian market and is suitable for milk collection centers where there is a big workload and fast and cost effective analysis is needed.
Key features:
- High capacity/ measurement speed: 90 samples per hour
- Model adapted to the Indian market and hot environmental conditions In-built cooling system
- Simple in operation, maintenance, calibration and installation
- Reliability and durability
- High accuracy and stability
- Portable and compact design
- Very small quantity of milk required
- Low power consumption
- No use of hazardous chemicals
- Connection to serial port of PC
- Software for PC supplied in the set
- Measurements stack - 256 last measurements are kept in the analyzer's memory
- Correction of calibration without computer
- Self-calibration on computer
- Receiving of the results of the measurements directly on the computer on-line and reading of the measurements stack
Electrical Parameters
AC Power Supply voltage - 100-260V, 50 Hz DC Power Supply voltage - 12V
Environmental Conditions
Ambient Temperature: +(10-40) °C Relative Air Humidity: 80% at 25 0С Atmospheric Pressure: 84 -107 Hecto Pascals
Mechanical Paramet
The analyzer's dimensions (H*W*D): 332 x 95 x 240 mm Weight: 3 kg Milk sample temperature: < + 35 °C Volume of milk sample per one measurement: 25 ml
Technical Description
Performance data
For samples in the temperature span +30 - 35 °C:
| Fat,% | SNF,% | |
Range of Measurement | 0 - 5 | 5 - 10 | 6 - 12 |
Allowed limits of the systematic component of the absolute error, % | +-0,05 | +-0,1 | +-0,1 |
Allowed limits of the random component of the absolute error, % | +-0,02 | +-0,03 | +-0,03 |
Performance data
For samples in the temperature span 15 - 40 °C:
| Fat,% | SNF,% | |
Range of Measurement | 0,5 - 5 | 5 - 10 | 6 - 12 |
Allowed limits of the systematic component of the absolute error, % | +-0,01 | +-0,15 | +-0,15 |
Allowed limits of the random component of the absolute error, % | +-0,03 | +-0,03 | +-0,03 |